1. Measuring Range and Accuracy of the Instrument:
1) Capacitance Measurement
a. Capacitance measuring range: 0.1μF to 3,300μF
b. Reactive power measuring range: 5 to 20,000 kvar
c. Measurement accuracy: ±(1.0% reading + 0.02μF)
d. Resolution: 0.001μF
2) Current Measurement
a. Current measuring range: 0 to 199.9mA; 0 to 1.999A; 0 to 19.99A; 0 to 199.9A; 0 to 1000A
b. Measurement accuracy: ±(3.0% reading + 0.05A)
c. Resolution: 0.01A
3) Inductance Measurement
a. Inductance measuring range: small inductance mode 0.1mH to 5mH
large inductance mode 5mH to 50H
b. Inductive reactance measuring range: 50mΩ to 20KΩ
c. Measurement accuracy: ±(3.0% reading + 0.05mH)
d. Resolution: 0.01mH
4) Resistance Measurement Range:
a. Resistance measuring range: small resistance mode 50mΩ to 1Ω
large resistance mode 1Ω to 20KΩ
b. Measurement accuracy: ±(3.0% reading + 0.05Ω)
c. Resolution: 0.01 Ω
2. Power Supply
1) Rated voltage: power frequency 220V±10%
2) Rated frequency: 50Hz;
3) Rated output: 2V/20V/500VA
3. Operating Conditions:
1) Ambient temperature: -10°C to +50°C
2) Relative humidity: ≤90%
4. Printing Method: LCD full Chinese character display
panel-mounted high-speed printer
5. Dimensions and Weight: 370×260×220mm/12kg