GKC-F High Voltage Switch Dynamic Characteristics Tester Uses * Measure the volt-ampere characteristics of linear resistive elements;
* Measure the forward and reverse volt-ampere characteristics of ordinary diodes;
* Measure the forward and reverse volt-ampere characteristics of Zener diodes;
* Measure the volt-ampere characteristics of small bulbs.
Technical Parameters and Features
* DC regulated power supply: Two ranges available (2V and 20V), continuously adjustable, with 3.5-digit digital display;
* Voltage display: Range 0-200V, 3.5-digit digital display;
* Current display: Four ranges available (200uA, 2mA, 20mA, 200mA), with 3.5-digit digital display;
* All use national standard connectors for manual wiring, which can cultivate hands-on ability.
GKC-F High Voltage Switch Dynamic Characteristics Tester Also known as:
High Voltage Switch Dynamic Characteristics Tester, Switch Mechanical Characteristics Tester, High Voltage Switch Comprehensive Tester, Circuit Breaker Dynamic Characteristics Tester, Circuit Breaker Mechanical Characteristics Tester, Circuit Breaker Comprehensive Tester, High Voltage Switch Timing Characteristics Tester, High Voltage Switch Characteristics Tester, Switch Characteristics Tester, Circuit Breaker Tester, Switch Mechanical Characteristics Tester,
Main Uses: Suitable for mechanical characteristics testing of high-voltage circuit breakers from imported and domestic manufacturers, including multi-oil, low-oil, SF6, vacuum, etc. Also suitable for time testing of electromagnetic action devices (such as contactors, relays, etc.).
Main Features: The instrument has a built-in DC operating power supply, and can flexibly use an external operating power supply or external synchronous trigger. Large-screen LCD display, with preset typical switch speed definitions; can switch to time waveform diagram to display 1 channel of close/open command, 12 channels of time transitions; S-T graph displays close/open motion curves, with dynamic analysis and calculation; V-S curve for intuitive observation of motion speed; I-T curve for observing the electromagnetic action process. Microcomputer printer prints data for retention. The instrument can save 255 test records, and can communicate with a computer; data management software can analyze, calibrate, and print test reports. The accessory box is equipped with sensors such as 0.1mm, 1mm linear, 1-degree angle photoelectric auxiliary contacts, as well as test leads, speed measurement mounting brackets, and other complete accessories.