Loop Resistance Tester with Printing
Precautions
| Fault Phenomenon | Cause Analysis | Troubleshooting | Remarks |
| No display upon power-on | 1) Power not connected | Connect the power supply | When replacing the fuse, use the same type; do not substitute with other types. |
| 2) The instrument's 5A fuse is not installed properly or is open | Reinstall the fuse or replace it | ||
| Output current cannot reach 200A | 1) Test leads too long, i.e., wire cross-section too small | Replace with wires of larger cross-section | |
| No current output | 1) The device under test is open circuit | Check the device Troubleshoot |
|
| Open circuit fault in the test loop | Check the test loop and clear the open circuit fault |
1. Before turning on the power switch, ensure the test button is released (not in test state) and the current adjustment knob is turned counterclockwise to the "zero" position.
2. The instrument should be placed indoors that is dry, ventilated, and free from corrosive gases.
3. After testing, wait until the power is fully discharged before removing test leads and other devices.
Transport and Storage
1. Transport
When the equipment needs to be transported, it is recommended to use our company's instrument packaging wooden case and shock-absorbing materials to avoid unnecessary damage during transit and prevent unnecessary losses.
When the equipment is transported without a wooden case, stacking is not allowed. When using our company's packaging case, the maximum stacking layers allowed is two.
During transportation, the instrument panel should face upward.
2. Storage
The equipment should be placed indoors that is dry, dust-free, ventilated, and free from corrosive gases. Without wooden case packaging, stacking is not allowed.
When storing, the panel should face upward. Place moisture-proof materials under the equipment to prevent dampness.