The fully automatic capacitance and inductance tester adopts synchronous sampling technology for standard capacitors and capacitors under test, which is not affected by power supply voltage fluctuations. In addition, the measurement process is fully automatic, avoiding errors caused by manual operation, thus featuring good stability, repeatability, accuracy, and reliability.



1. Capacitance range: 0.2μF~2,000μF;

Capacity range: 5~20,000 kvar;

Measurement accuracy: ±1.5%×reading±2 digits;

Resolution: ±1.0%rdg;

2. Inductance range: 1mH~9.99H; Measurement accuracy: ±1.5%±2 digits

3. Output measurement voltage: AC 26V/500VA; 50Hz;

4. Display mode: large-screen LCD with full Chinese output, TPμp-40 panel-type thermal printer

5. Dimensions and weight: 370×370×220 mm/ 8 kg

6. Working conditions:

Ambient temperature: 0℃ ~+40℃, relative humidity: ≤90%

Power supply: AC 220V±10%; 50Hz;

Measurement range, resolution, and error values



























Capacitance measurement range: Error value:
0.020μF ~ 0.200μF ±0.5%
0.200μF ~ 2.000μF ±0.5%
2.000μF ~ 20.00μF ±0.5%
20.00μF ~ 200.0μF ±0.5%
200.0μF ~ 2000.μF ±0.5%





























Capacitor reactive power 0 ~ 20.00Mvar Error value ±1%
Capacitor active power 0 ~ 20.00kW Error value ±1%
Capacitor dissipation factor 0 ~ 20.00% Error value ±1%
Capacitor resistance component 0 ~ 10.00MΩ Error value ±1%

























Inductance measurement range Error value
0.10mH ~ 0.200H ±0.5%
2.000H ~ 20.00H ±0.5%
20.00H ~ 200.0H ±0.5%
200.0H ~ 2000.H ±0.5%





























Inductor reactive power 0 ~ 20.00Mvar Error value ±1%
Inductor active power 0 ~ 20.00kW Error value ±1%
Inductor dissipation factor 0 ~ 20.00% Error value ±1%
Inductor resistance component 0 ~ 10.00MΩ Error value ±1%