Fault Analysis











































Fault Phenomenon



Cause Analysis



Troubleshooting Method



Remarks



No display upon power-on



1) Power not connected



Connect the power supply



When replacing the fuse, use the same model; do not substitute with other types.



2) The instrument's 5A fuse is not installed properly or is open



Reinstall the fuse or replace it


   

Output current cannot reach 200A



1) Test line is too long, i.e., conductor cross-section is too small



Replace with a conductor of larger cross-section


 

No current output



1) The device under test is open circuit



Check the device



Eliminate the fault


 

2) There is an open circuit fault in the test loop



Check the test loop and eliminate the open circuit fault


   


Precautions



1. Before turning on the power switch, ensure the test button is released (non-test state) and the current adjustment knob is turned counterclockwise to the "zero" position.



2. The instrument should be placed in a dry, ventilated indoor area free from corrosive gases.



3. After testing, wait until the power is completely discharged before removing test leads and other equipment.



Transportation and Storage



1. Transportation



When the equipment needs to be transported, it is recommended to use the company's wooden packaging box and shock-absorbing materials to avoid unnecessary damage during transit, which could cause you unnecessary losses.



When the equipment is transported without a wooden box, stacking is not allowed. When using the company's packaging box, the maximum stacking layers allowed is two.



During transportation, the instrument panel should face upward.



2. Storage



The equipment should be placed in a dry, dust-free, ventilated indoor area free from corrosive gases. Without wooden box packaging, stacking is not allowed.



When storing, the panel should face upward. Place moisture-proof materials at the bottom of the equipment to prevent moisture damage.